Model WDC-80TD Low temperature Chamber for DWTT Impact Specimen

2024-12-24 16:07 havePeople browsing 小编

classificationImpact Test Machine

productModel WDC-80TD Low temperature Chamber for DWTT Impact Specimen

size

length

price

phone 13969111546

Applications:Model WDC-80TD low chamber is newly designed according to Charpy no...

Details

Applications:

Model WDC-80TD low chamber is newly designed according to Charpy notch Impact Test Method for Metal Material, and adopts alcohol cooling technology. It utilizes the heat balance principle and cycle stirring method to realize the constant temperature cooling to impact specimen with the reliable performance. Digital display, automatic control temperature, and it is the ideal equipment for cooling and keeping temperature to impact specimen for WDTT drop weight as ASTM E436 and API 5L standards.

 

Main Specifications

Temperature control range30~-80

Constant temperature accuracy: ≤±0.5ºC

Temperature dropping speed:  Within 80 minutes the temperature will drop to -80 from Ambient

Dimension: 1515×1010×1120mm

Effective working space: 600 ×400 ×200mm

Specimen quantity more than 24 pcs

Specimen dimension350x75mm (Length x width), the thickness will be as actual material thickness)

Cooling media: Alcohol or other non-freezing solution (60L)

Power supply: 220V ~ 240V50Hz14kW